SAMPLE COLLECTION: Saliva collection from all the selected subjects will be performed following the Salivette (SARSTEDT) manufacturer's instructions. To limit variability in salivary content not related to ALS, saliva will be obtained from all subjects at a fixed time, after an appropriate lag time from feeding and teeth brushing. Pre-analytical parameters (i.e. storage temperature and time between collection and processing), dietary and smoking habit will be properly recorded. Briefly, the swab will be removed, placed in the mouth and chewed for 60 seconds to stimulate salivation. Then the swab will be centrifuged for 2 minutes at 1,000 g to remove cells fragments and food debris. Collected samples will be stored at -80° C.
SAMPLE PROCESSING: Before the Raman acquisition, saliva samples will be filtered with different cut-off ranges (3, 10 and 30 kDa), collecting and analysing by RS the eluted sample and discarding the concentrated counterpart. For the Raman analysis, a drop of each sample will be casted on an aluminium foil in order to achieve the Surface Enhanced Raman Scattering (SERS).
DATA COLLECTION: Raman and SERS spectra will be acquired using an Aramis Raman microscope (Horiba Jobin-Yvon, France) equipped with a laser light source operating at 785 nm with laser power ranging from 25-100% (Max power 512 mW). Acquisition time between 10-30 seconds will be used. The instrument will be calibrated before each analysis using the reference band of silicon at 520.7 cm-1. Raman spectra will be collected from 15 points following a line-map from the edge to the centre of the drop. Spectra will be acquired in the region between 400 and 1800 cm-1 using a 50x objective (Olympus, Japan). Spectra resolution is about 1.2 cm-1. The software package LabSpec 6 (Horiba Jobin-Yvon, France) will be used for map design and the acquisition of spectra.
DATA PROCESSING: All the acquired spectra will be fit with a fifth-degree polynomial baseline and normalized by unit vector using the dedicated software LabSpec 6. The contribution of the substrate will be removed from each spectra. The statistical analysis to validate the method, will be performed using a multivariate analysis approach. Principal Component analysis (PCA) will be performed in order to reduce data dimensions and to evidence major trends. The first 20 resultant Principal Components (PCs) will be used in a classification model, Linear Discriminant Analysis (LDA), to discriminate the data maximizing the variance between the selected groups. The smallest number of PCs will be selected to prevent data overfitting. Leave-one-out cross-validation and confusion matrix test will be used to evaluate the method sensitivity, precision and accuracy of the LDA model. Mann-Whitney will be performed on PCs scores to verify the differences statistically relevant between the analysed groups. Correlation and partial correlation analysis will be performed using the Spearman's test, assuming as valid correlation only the coefficients with a p-value lower than 0.05. The statistical analysis will be performed using Origin2018 (OriginLab, USA).