In this study, the level of oxygen within the blood will be reduced in a controlled manner by reducing the concentration of oxygen the participant breathes. The accuracy of a noninvasive pulse oximeter sensor will be assessed by comparison to the oxygen saturation measurements from a laboratory blood gas analyzer.
It is required that the Accuracy Root Mean Square (ARMS) performance of the Stryker pulse oximetry sensors will meet a specification of +/-3% or better in non-motion conditions for the range of 70 - 100% SaO2 (typically, saturation is determined once with air breathing and then at one of 3 levels, e.g. 95%, 85%, and 75% saturation for about 30-60 seconds at each level), thereby demonstrating an acceptable SpO2 accuracy performance specification.