This study is designed to evaluate the performance of the NM Meter when used by CLIAW operators in terms of both accuracy to an established reference method and precision. CLIAW operators will receive no training or prompting on how to use the NM Creatinine and eGFR Meter, operating only using the instructions found in printed labeling materials (IFU, QRG, packaging).
In determining the accuracy of the device, subjects will be tested using the NM Creatinine and eGFR Meter alongside a central laboratory reference method and the results will be compared to one another. This portion of the study is referred to as Method Comparison.
In determining the precision of the device, multiple creatinine determinations will be made with the NM Creatinine and eGFR meter using both stabilized control materials (testing over the course of 20 days) and venous whole blood specimens (tested over the course of a single day). The difference between the multiple test results will determine the device's level of precision. This portion of the study is referred to as Precision.