Death and vascular tears occur in up to 1% of the patients undergoing lead extraction procedures. These complications are due to abnormal force vectors during the procedure. The investigators postulate that use of bidirectional traction (superior and inferior directions) and feedback from a digital force gauge is associated with minimal lead displacement compared to conventional lead extraction. Lead displacement is an indirect risk factor for potentially fatal lead extraction complications
Use of force no greater than 8 pounds of tension will improve lead extraction and minimize lead displacement compared to current standard of care methods which do not limit the force applied. Among subjects undergoing lead extraction procedures for CIED infection, bidirectional rail (superior and inferior) and traction force objective feedback using Force gauge is associated with minimal displacement (\<1cm) of the leads (surrogate marker) on fluoroscopy compared to conventional lead extraction. • The research nature of the protocol is explicitly explained to the patient and an informed consent is explained obtained from the patient prior to the procedure. Only patients with CIED infections requiring lead extraction are enrolled in the study. Patients are risk stratified based on our current risk stratifications scheme. Multiple superior and inferior venous accesses are obtained during the procedure. The leads are prepared using locking stylets. Baseline fluoroscopic image of the position of the leads is stored. A deflectable sheath and traction forceps are deployed from the groin to provide downward rail in the force feedback arm. Locking stylet from the lead is in turn attached to a Force gauge (McMaster-Carr Santa Fe Springs, CA). The procedure is performed using continuous fluoroscopic image save tool during the procedure for post procedural viewing and analysis. The patients are randomized to the active arm (to keep force around 8 lb) or the control arm (tactile force discretion based on the operator. The investigator is blinded to the force gauge reading for half of the patients enrolled in the study. For the other half of the patients, the operator optimizes the force to prevent overexertion (\<8-9 pounds). Three point along the lead course evaluated during the study. Position of downward turn of the lead in the SVC, heel of the lead (only in case of RV lead) and the lead tip positions are evaluated. The investigators postulate that there will be at least 5 cine-fluoroscopic views per every lead. The investigators hypothesize that use of bidirectional rail and force feedback is associated with minimal displacement of the lead on fluoroscopy.
Age
18 - No limit years
Sex
ALL
Healthy Volunteers
No
Mayo Clinic in Rochester
Rochester, Minnesota, United States
Start Date
December 20, 2018
Primary Completion Date
December 10, 2020
Completion Date
December 10, 2020
Last Updated
June 23, 2022
22
ACTUAL participants
Force gauge
OTHER
Lead Sponsor
Mayo Clinic
Data Source & Attribution
This clinical trial information is sourced from ClinicalTrials.gov, a service of the U.S. National Institutes of Health.
Modifications: This data has been reformatted for display purposes. Eligibility criteria have been parsed into inclusion/exclusion sections. Location data has been geocoded to enable distance-based search. For the authoritative and most current information, please visit ClinicalTrials.gov.
Neither the United States Government nor Clareo Health make any warranties regarding the data. Check ClinicalTrials.gov frequently for updates.
View ClinicalTrials.gov Terms and Conditions