You may be eligible if
Age greater than or equal to 18 years
Able and willing to complete MRI testing
Able to provide written informed consent
Available for follow-up visit at the study site
Implanted with an ICD System consisting only of an Iforia DR-T and Linoxsmart S 65 / Linoxsmart SD 65/18 with Setrox S 53 atrial lead or Iforia VR-T DX and Linoxsmart S DX 65/15 or Linoxsmart S DX 65/17.
ICD implanted pectorally
All ICD system components implanted, repositioned, or exchanged at least 5 weeks prior to enrollment
Underlying rhythm identifiable during sensing test
All pacing thresholds are measureable and are less than or equal to 2.0 V @ 0.4 ms (Criterion does not apply to atrial pacing threshold for subjects with an Iforia DR-T that have AF and are permanently programmed to VVI and for subjects with an Iforia VR-T DX)
Ventricular sensing is measurable and \>6.5mV
Absence of phrenic nerve stimulation at 5.0V and 1.0 ms
Pacing impedance is between 200 and 1500 ohm
Shock Impedance is between 30 and 90 ohm
Able and willing to use the CardioMessenger II
At the pre-MRI procedure, the following procedure-related criteria must be met for the subject to undergo the MRI scan:
Absence of phrenic nerve stimulation at 5.0V @ 1.0 ms
\|pacing threshold at Pre-MRI follow-up - pacing threshold at baseline\| ≤ 0.5 V (ICD Lead Only)
All pacing thresholds are measureable and are less than or equal to 2.0 V @ 0.4 ms (Criterion does not apply to atrial pacing threshold for subjects with an Iforia DR-T that have AF and are permanently programmed to VVI and for subjects with an Iforia VR-T DX.)
Ventricular sensing is measurable and \>6.5mV
The ICD system has been implanted for at least six weeks.
Subject did not have an ICD or lead explant, exchange or reposition in the previous 6 weeks.
All lead pacing impedances are between 200 and 1500 ohm