Finding studies
Finding studies
Take this into the appointment.
Saves the questions and what to expect into your notes, next to the visit they belong to.
Lead
Legacy Health System
With
The objective of the current human study is to verify the components of the Artificial Pancreas Control system during an inpatient study. This master controller software is designed to be used in conjunction with two subcutaneous continuous glucose monitoring systems to regulate blood glucose levels as well as two Omnipod pumps, one for administering insulin and one for administering glucagon. The sensors communicate wirelessly with two sensor receivers which will be interfaced with the APC by wireless USB connection. The insulin and glucagon pumps will be controlled by the APC through a wireless USB connection. The algorithm included in the APC is an automated version of the Adaptive Proportional Derivative (APD) insulin and glucagon control algorithm, which was previously studied as an investigational device. The APD has been studied in vivo (in 28 experiments, each 33 hr in length, with manual adjustment of pumps) and no serious adverse effects were noted. Manual input of the glucose sensor data and insulin/glucagon infusion rates will no longer be necessary. The APC will be tested in vivo during 28 hour experiments in an inpatient setting in preparation for outpatient testing.
Age
21–65
Sex
ALL
Healthy volunteers
Not accepted