According to the "European Society of Hypertension International Protocol Revision 2010 for the Validation of Blood Pressure Measuring Devices in Adults", 33 patients have their blood pressure measured by two observers using parallel connected mercury sphygmomanometers, blinded from each other's result, and then by the supervisor using the tested device, NISSEI DM 3000. The measurements using the mercury sphygmomanometers are repeated 5 times and in between them, 4 measurements using the tested device take place. If the results of the comparison of these measurements according to the analysis required by the protocol, meet the required criteria, the device is regarded as "Pass", otherwise as "Fail".
In this study, instead of consecutive, simultaneous measurements of the two mercury sphygmomanometers and tha tested device will be taking place. This will be so to exclude any inaccuracy due to blood pressure variability and because the tested device has a controllable deflation rate and allows simultaneous measurements with the mercury sphygmomanometers. In addition, the investigators will be performing 6 set of measurements instead of 4.
Age
25–any
Sex
ALL
Healthy volunteers
Accepted
You may be eligible if
Above 25 years of age
Sinus rhythm
Informed consent
You may not be if
Sustained arrhythmia
Pregnancy
Poor quality Korotkoff sounds
Unavailable cuff size
Completed blood pressure range (according to protocol)