Acute Feasibility Investigation of a New S-ICD Electrode Arrangement for Reduction of Defibrillation Energies
This is an early-stage feasibility study. This study seeks to characterize the performance of the new electrode shock configuration with the S-ICD
Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD, EMBLEM®, Boston Scientific Corp) delivers 80 J shocks from an 8 cm left-parasternal coil (LPC) to a 59 cc left lateral pulse generator (PG). A system that defibrillates with lower energy could significantly reduce PG size. Computer modeling and animal studies suggest a 2nd shock coil either parallel (P) to LPC or transverse (T) from xiphoid to PG pocket would significantly reduce shock impedance and defibrillation threshold (DFT). Objective: Acutely test defibrillation efficacy of P and T configurations in patients receiving an S-ICD. Methods: Testing was performed in patients receiving a conventional S-ICD system (C) for standard indications. Patients at risk for conversion failure, stroke or infection were excluded. A 65 J success with the conventional S-ICD was required prior to investigational testing. A second electrode was temporarily inserted from the xiphoid incision using the same introducer and sheath tools and connected to the PG with an investigational Y-adapter (Model 3598, Boston Scientific). Phase 1 (n=11) tested only P at a starting energy of 50 J with subsequent tests at 65 or 30 J then 40 or 20 J depending upon outcome. Phase 2 (n=21) tested both P and T in random order, starting at 30 J then either 40 or 20 J.
Age
18 - No limit years
Sex
ALL
Healthy Volunteers
No
Academisch Medisch Centrum
Amsterdam, Netherlands
St. Antonius Ziekenhuis
Nieuwegein, Netherlands
Erasmus MC - University Medical Center Rotterdam
Rotterdam, Netherlands
Isala Klinieken
Zwolle, Netherlands
Start Date
November 29, 2018
Primary Completion Date
April 1, 2022
Completion Date
April 1, 2022
Last Updated
May 9, 2024
42
ACTUAL participants
commercially released subcutaneous implantable cardioverter defibrillator (S-ICD) and the Investigational S-ICD Adapter.
DEVICE
Lead Sponsor
Boston Scientific Corporation
Data Source & Attribution
This clinical trial information is sourced from ClinicalTrials.gov, a service of the U.S. National Institutes of Health.
Modifications: This data has been reformatted for display purposes. Eligibility criteria have been parsed into inclusion/exclusion sections. Location data has been geocoded to enable distance-based search. For the authoritative and most current information, please visit ClinicalTrials.gov.
Neither the United States Government nor Clareo Health make any warranties regarding the data. Check ClinicalTrials.gov frequently for updates.
View ClinicalTrials.gov Terms and Conditions